60 Watt Medical (MVLT) Features l l l l l l l 4 x 2 x 1.07 inches Ultra high efficiency > 82% 60 W from 1 to 3 outputs Low leakage current < 250 A Nemko, UL & CSA approvals to IEC60601 EN55022-B, CISPR22-B, FCC Part15 Level B, IEC60601-1-2 Cover kit accessory available Electrical Specifications AC Input Input Frequency Input Current Inrush Current Leakage Current Efficiency Hold-up Time Output Power1 Line Regulation Load Regulation Transient Response Rise Time Set Point Tolerance Over Current Protection Over Voltage Protection (For V1 & triple output model only) Short Circuit Protection Switching Frequency Operating Temperature Storage Temperature Relative Humidity Altitude MTBF Isolation Voltage Cooling 90-132 V & 180-264 V, Auto Ranging 47-63 Hz 120 VAC: 1.25 A max. 230 VAC: 0.65A max. 120 VAC: 40 A max. 230 VAC: 75 A max. 120V AC: < 140 A 230 VAC: < 250 A 120 VAC: 80% typical 230 VAC: 80% typical 120 VAC: 6 ms 230 VAC: 6 ms 50 to 60 W +/-0.3% V1: +/-0.5%; V2 & V3: +/-5% < 10%, 50% to 100% load change, 50/60 Hz, 50% duty cycle, 0.1 A/s, recovery time < 5 ms < 100 ms V1: +/-3%; V2 & V3: +/-5% 130% typical above rating 3.95 V +/-0.25 V for 3.3 V; 6.2 V +/-0.4 V for 5 V Short term, autorecovery Boost converter: 105 kHz typical Resonant converter: 55 kHz typical 0 to 70C, refer derating curve -40 to +85C 95% Rh, noncondensing Operating: 10,000 ft.; Nonoperating: 40,000 ft. 0.91m Hours, Telcordia SR332 Issue-3 Min. 5700 VDC between input to output Convection 41 4EM-17-002 39-DE60-43348-002 / B1 Model Number Voltage LFMVLT60-1200 V1=5.1 V LFMVLT60-1201 V1=12 V LFMVLT60-1202 V1=15 V LFMVLT60-1203 V1=24 V LFMVLT60-1204 V1=48 V LFMVLT60-3200 V1=5.2 V, V2=12.5 V, V3=-12.8 V LFMVLT60-3201 V1=5.2 V, V2=24 V, V3=-12.8 V LFMVLT60-3202 V1=5.2 V, V2=14.6 V, V3=-14.8 V LFMVLT60-3203 V1=3.3 V, V2=5.2 V, V3=-12.8 V LFVLT60-CK metal cover kit accessory Max. Load2 Min. Load Ripple3 10.0 A 5.0 A 4.0 A 2.5 A 1.25 A V1=8.0 A, V2=3.0 A, V3=0.5 A V1=8.0 A, V2=1.5 A, V3=0.5 A V1=8.0 A, V2=2.5 A, V3=0.5 A V1=8.0 A, V2=3.0 A, V3=0.5 A 0.2 A 0.1A 0.1 A 0.1 A 0.05 A V1=0.5 A, V2=0.1 A, V3=0.0 A V1=0.5 A, V2=0.1 A, V3=0.0 A V1=0.5 A, V2=0.1 A, V3=0.0 A V1=1.0 A, V2=0.1 A, V3=0.0 A 1% 1% 1% 1% 1% 1% 1% 1% V1=1.5%, V2 & V3=1% Connectors J1 Pin 1 Pin 2 Spade Connector J2 Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 AC NEUTRAL AC LINE EARTH V1 V1 RTN RTN V3 V2 Notes 1. Rating of MVLT60-1200 is 50 W and MVLT60-3203 is 45 W. 2. Maximum current per output channel. Do not exceed total output power rating. 3. Ripple is peak to peak with 20 MHz bandwidth and 10 F (Tantalum capacitor) in parallel with a 0.1 F capacitor at rated line voltage and load ranges. 4. Specifications are for nominal input voltage, 25C and max. load unless otherwise stated. 5. Derate output power linearly to 80% from 90 VAC to 80 VAC input. 2 4EM-17-002 39-DE60-43348-002 / B1 Innovations in Power Mechanical Specifications AC Input Connector (J1) EARTH DC Output Connector (J2) Dimensions Weight Molex: 26-60-4030 or equivalent Mating: 09-50-3031; Pins: 08-50-0106 Molex: 19705-4301 or equivalent Mating: 190030001 Tyco: 640445-6 or equivalent Mating: 647402-6; Pins: 3-647409-1 4.0 x 2.04 x 1.07 inches (101.6 x 51.82 x 27.18 mm) 150 g EMC CE Mark Conducted Emissions Static Discharge RF Field Susceptibility Fast Transients/Bursts Radiated Emissions Surge Susceptibility Complies with LVD Directive EN55022-B, CISPR22-B, FCC PART15-B EN61000-4-2, Level-3 EN61000-4-3, Level-3 EN61000-4-4, Level-3 EN55022-B, CISPR22-B, FCC PART15-B To be controlled in end system EN61000-4-5, Level-3 Safety Safety Standard(s) Approval Agency Safety File Number(s) EN60601-1, IEC60601-1 (ed.3), UL60601-1 (1st Edition), CSA C22.2 No. 601.1, Class 1 SELV Nemko, UL, C-UL Nemko: P11214761 UL: E173812 43 4EM-17-002 39-DE60-43348-002 / B1 Derating Curve Mechanical Drawing Notes: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following 1. Stand off, used to mount PCB has OD of 5.4 mm max. 2. Screws, used to fix PCB on stand off, have head dia of 6.0 mm max. 3. Washer, if used, to have dia of 6.5 mm max. 4 Innovations in Power 4EM-17-002 39-DE60-43348-002 / B1